Guidance and manufacture’s declaration – electromagnetic immunity –
for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration – electromagnetic immunity
The
pro6
+
is intended for use in the electromagnetic environment specied below. The customer or user of the
pro6
+
should assure that it is used in such an environment.
Immunity test IEC 60601 test Compliance Electromagnetic environment
level level - guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the
pro6
+
, including cables, than
the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
Conducted RF 3 Vrms 3 Vrms
IEC 61000-4-6 150 kHz to 80 MHz
Radiated RF 3 V/m 3 Vrms 80 MHz to 800 MHz
IEC 61000-4-3 80 MHz to 2.5 GHz 800 MHz to 2.5 GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according
to the transmitter manufacturer and d is the
recommended separation distance in metres (m).
Field strengths from xed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption
and reection from structures, objects and people.
a
Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to xed RF transmitters, an electromagnetic site
survey should be considered. If the measured eld strength in the location in which the
pro6
+
is used exceeds
the applicable RF compliance level above, the
pro6
+
should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the
pro6
+
.
b
Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3 V/m.
Pd 167.1=
Pd 167.1=
Pd 167.1=
12